Titel: Compressive effect and collapse behavior of three different transsacral implants in sacral fragility fractures : a retrospective analysis of 106 cases
Autor(en): Fischer, ChristianIn der Gemeinsamen Normdatei der DNB nachschlagen
Klauke, Friederike
Schenk, PhilippIn der Gemeinsamen Normdatei der DNB nachschlagen
Bauerfeld, Henning
Kobbe, PhilippIn der Gemeinsamen Normdatei der DNB nachschlagen
Mendel, ThomasIn der Gemeinsamen Normdatei der DNB nachschlagen
Erscheinungsdatum: 2024
Art: Artikel
Sprache: Englisch
Zusammenfassung: Purpose: The aim of this study were the retrospective evaluation of the compressive effect and complication rates of transsacral stabilization of osteoporosis-associated sacral fragility fractures in 106 patients using three different implants (6.0 mm sacral bar, n = 32; 7.3 mm screw, n = 26; 7.5 mm ISG-Rod System, n = 48) with regard to the image morphological and clinical-perioperative outcome. Methods: For this purpose, the sacral width was determined preoperatively and postoperatively using multiplanar CT reconstructions and correlated with the measured bone density (HU). The results were compared with each other on an implant-specific basis. Results: A significant compressive effect was found for all implants (6.0 mm sacral bar 7.1 ± 3.4 mm, 7.3 mm screw 6.9 ± 1.8 mm, 7.5 mm ISG-Rod System 8 ± 2.4 mm). No implant-specific difference in compression could be detected. Overall, the washers broke into the iliac cortex in 9% of cases. The subgroups did not differ significantly in this respect (6.0 mm sacral bar: 4 [13%], 7.3 mm screw 1 [1%], 7.5 mm ISG-Rod System (5 [10%], p = 0.581). A correlation between the degree of osteoporosis and the compressive effect could not be demonstrated. Significant implant-specific differences were found in the incision-suture time, with only ø0:39 ± 0:13 h required for implantation of the 7.5 mm ISG Rod System (6.0 mm sacral bar: ø1:09 ± 0:22 h, 7.3 mm screw: ø0:55 ± 0:20 h). The fluoroscopy time was significantly lower with the 7.3 mm screw (ø0:57 ± 0:23 min) and the 7.5 mm ISG Rod System (ø0:42 ± 00:17 min) than with the 6.0 mm sacral bar (ø1:36 ± 0:46 min). Conclusion: A significant compressive effect was demonstrated with all three implants. No implant-specific complications or surgical site complications were identified in either the overall cohort or the subgroups. The 7.5 mm ISG Rod System shows advantages with regard to the duration of surgery and fluoroscopy.
URI: https://opendata.uni-halle.de//handle/1981185920/119889
http://dx.doi.org/10.25673/117929
Open-Access: Open-Access-Publikation
Nutzungslizenz: (CC BY-NC-ND 4.0) Creative Commons Namensnennung - Nicht kommerziell - Keine Bearbeitungen 4.0 International(CC BY-NC-ND 4.0) Creative Commons Namensnennung - Nicht kommerziell - Keine Bearbeitungen 4.0 International
Journal Titel: European journal of trauma and emergency surgery
Verlag: Springer Medizin
Verlagsort: Heidelberg
Band: 50
Heft: 6
Originalveröffentlichung: 10.1007/s00068-024-02629-1
Seitenanfang: 2963
Seitenende: 2973
Enthalten in den Sammlungen:Open Access Publikationen der MLU

Dateien zu dieser Ressource:
Datei Beschreibung GrößeFormat 
s00068-024-02629-1.pdf1.61 MBAdobe PDFMiniaturbild
Öffnen/Anzeigen