Bitte benutzen Sie diese Kennung, um auf die Ressource zu verweisen: http://dx.doi.org/10.25673/118392
Titel: Comparative investigation of beam quality index Q on flattening filter and flattening filter free 6 and 15 MV photon beams of an medical linear accelerator
Autor(en): Schaks, MichaelIn der Gemeinsamen Normdatei der DNB nachschlagen
Zähle, Johannes
Janich, MartinIn der Gemeinsamen Normdatei der DNB nachschlagen
Hübsch, Patrick
Homonnay, Nico
Gerlach, Reinhard
Erscheinungsdatum: 2017
Art: Artikel
Sprache: Englisch
Zusammenfassung: An Elekta Synergy LINACwas used to investigate the beam quality index Q and the beam quality correction factor kQ,which is used for absolute dosimetry, in off-axis photon fields. It was foundthat the beam quality index Q of the photonenergies with flattening filter decreases with increasing distance to the central axis, at 6 MV by 4.27%and at 15 MVby3.98% inside a range of 15 cm off-axis. The beam quality index for flattening filter free photon fields also decreases with increasing distance to the central axis. In this case Q changed only by 1.01% inside the above range.
URI: https://opendata.uni-halle.de//handle/1981185920/120351
http://dx.doi.org/10.25673/118392
Open-Access: Open-Access-Publikation
Nutzungslizenz: (CC BY-NC-ND 4.0) Creative Commons Namensnennung - Nicht kommerziell - Keine Bearbeitungen 4.0 International(CC BY-NC-ND 4.0) Creative Commons Namensnennung - Nicht kommerziell - Keine Bearbeitungen 4.0 International
Journal Titel: Current directions in biomedical engineering
Verlag: De Gruyter
Verlagsort: Berlin
Band: 3
Heft: 2
Originalveröffentlichung: 10.1515/cdbme-2017-0059
Seitenanfang: 287
Seitenende: 290