Title: A retrospective European multicenter analysis of the functional outcomes after active middle ear implant surgery using the third generation vibroplasty couplers
Author(s): Rahne, Torsten
Skarzynski, Piotr Henryk
Hagen, Rudolf
Radeloff, Andreas DanielLook up in the Integrated Authority File of the German National Library
Lassaletta, Luis
Barbara, Maurizio
Plontke, Stefan K.-R.Look up in the Integrated Authority File of the German National Library
Mlynski, Robert ArndtLook up in the Integrated Authority File of the German National Library
Issue Date: 2021
Type: Article
Language: English
Abstract: Purpose: To evaluate the safety and performance of three novel vibroplasty couplers that allow attachment of the floating mass transducer of a transcutaneous active middle ear implant (AMEI) to the round window (RW) membrane, the long process (LP), or the incus body and the short process (SP) of the incus. Methods: Retrospective multicenter cohort study of 25 AMEI users with sensorineural or mixed hearing loss that were among the first implanted with an AMEI vibrating ossicular prosthesis in combination with the third generation of vibroplasty couplers between 2014 and 2016. Main Outcome Measures were bone-conduction pure-tone and vibroplasty thresholds, postoperative aided sound field thresholds and postoperative aided word recognition score (WRS). Results: Bone conduction threshold changes of more than 10 dB in 4PTABC were observed in two subjects. A mean improvement of 57.8% in speech recognition was observed with a mean WRS at 65 dB SPL improving from 14.8% (SD 21.9%) preoperatively to a mean aided score of 72.6% (SD 18.6%). Sound field thresholds improved from an average 4PTASF of 64.1 dB HL (SD 9.8 dB HL) to 37.0 dB HL (SD 8.9 dB HL), resulting in a mean functional gain of 27.1 dB. There was no significant difference in WRS or functional gain between the coupler types. Conclusion: Initial experience shows that all three third generation vibroplasty couplers represent safe and efficient attachment options for the FMT allowing the surgeon to choose the coupling type based on the present pathology.
URI: https://opendata.uni-halle.de//handle/1981185920/38511
http://dx.doi.org/10.25673/38268
Open Access: Open access publication
License: (CC BY 4.0) Creative Commons Attribution 4.0(CC BY 4.0) Creative Commons Attribution 4.0
Sponsor/Funder: Publikationsfond MLU
Journal Title: European archives of oto-rhino-laryngology and head & neck
Publisher: Springer
Publisher Place: Berlin
Volume: 278
Original Publication: 10.1007/s00405-020-06064-x
Page Start: 67
Page End: 75
Appears in Collections:Open Access Publikationen der MLU

Files in This Item:
File Description SizeFormat 
Rahne2021_Article_ARetrospectiveEuropeanMulticen.pdf682.47 kBAdobe PDFThumbnail
View/Open