Bitte benutzen Sie diese Kennung, um auf die Ressource zu verweisen:
http://dx.doi.org/10.25673/118392| Titel: | Comparative investigation of beam quality index Q on flattening filter and flattening filter free 6 and 15 MV photon beams of an medical linear accelerator |
| Autor(en): | Schaks, Michael Zähle, Johannes Janich, Martin Hübsch, Patrick Homonnay, Nico Gerlach, Reinhard |
| Erscheinungsdatum: | 2017 |
| Art: | Artikel |
| Sprache: | Englisch |
| Zusammenfassung: | An Elekta Synergy LINACwas used to investigate the beam quality index Q and the beam quality correction factor kQ,which is used for absolute dosimetry, in off-axis photon fields. It was foundthat the beam quality index Q of the photonenergies with flattening filter decreases with increasing distance to the central axis, at 6 MV by 4.27%and at 15 MVby3.98% inside a range of 15 cm off-axis. The beam quality index for flattening filter free photon fields also decreases with increasing distance to the central axis. In this case Q changed only by 1.01% inside the above range. |
| URI: | https://opendata.uni-halle.de//handle/1981185920/120351 http://dx.doi.org/10.25673/118392 |
| Open-Access: | Open-Access-Publikation |
| Nutzungslizenz: | (CC BY-NC-ND 4.0) Creative Commons Namensnennung - Nicht kommerziell - Keine Bearbeitungen 4.0 International |
| Journal Titel: | Current directions in biomedical engineering |
| Verlag: | De Gruyter |
| Verlagsort: | Berlin |
| Band: | 3 |
| Heft: | 2 |
| Originalveröffentlichung: | 10.1515/cdbme-2017-0059 |
| Seitenanfang: | 287 |
| Seitenende: | 290 |
| Enthalten in den Sammlungen: | Open Access Publikationen der MLU |
Dateien zu dieser Ressource:
| Datei | Beschreibung | Größe | Format | |
|---|---|---|---|---|
| 10.1515_cdbme-2017-0059.pdf | 549.97 kB | Adobe PDF | ![]() Öffnen/Anzeigen |
Open-Access-Publikation