Titel: Image fusion of high-resolution DynaCT and T2-weighted MRI for image-guided programming of dDBS
Autor(en): Al-Jaberi, Fadil
Moeskes, Matthias
Skalej, Martin
Fachet, MelanieIn der Gemeinsamen Normdatei der DNB nachschlagen
Hoeschen, ChristophIn der Gemeinsamen Normdatei der DNB nachschlagen
Erscheinungsdatum: 2025
Art: Artikel
Sprache: Englisch
Zusammenfassung: Objectives: This study aimed to develop a semi-automated registration method for aligning preoperative non-contrast T2-weighted MRI with postoperative high-resolution cone-beam CT (DynaCT) in patients undergoing directional deep brain stimulation (dDBS) surgery targeting the subthalamic nucleus (STN). The aim was to facilitate image-guided programming of DBS devices and postoperative verification of the alignment of segmented contacts. Materials and Methods: A dataset of ten patients undergoing bilateral dDBS implantation was retrospectively collected, including DynaCT (acquired postoperatively) and non-contrast T2-weighted MRI (obtained preoperatively). A semi-automated registration method was used, employing manual initialization due to dissimilar anatomical information between DynaCT and T2-weighted MRI. Image visualization, initial alignment using a centered transformation initializer, and single-resolution image registration involving the Simple Insight Toolkit (SimpleITK) library were performed. Manual landmark-based alignment based on anatomical landmarks and evaluation metrics such as Target Registration Error (TRE) assessed alignment accuracy. Results: The registration method successfully aligned all images. Quantitative evaluation revealed an average of the mean TRE of 1.48 mm across all subjects, indicating satisfactory alignment quality. Multiplanar reformations (MPRs) based on electrode-oriented normal vectors visualized segmented contacts for accurate electrode placement. Conclusions: The developed method demonstrated successful registration between preoperative non-contrast T2-weighted MRI and postoperative DynaCT, despite dissimilar anatomical information. This approach facilitates accurate alignment crucial for DBS programming and postoperative verification, potentially reducing the programming time of the DBS. The study underscores the importance of image quality, manual initialization and semi-automated registration methods for successful multimodal image registration in dDBS procedures targeting the STN.
URI: https://opendata.uni-halle.de//handle/1981185920/121321
Open-Access: Open-Access-Publikation
Nutzungslizenz: (CC BY 4.0) Creative Commons Namensnennung 4.0 International(CC BY 4.0) Creative Commons Namensnennung 4.0 International
Journal Titel: Brain Sciences
Verlag: MDPI AG
Verlagsort: Basel
Band: 15
Heft: 5
Originalveröffentlichung: 10.3390/brainsci15050521
Seitenanfang: 1
Seitenende: 12