Bitte benutzen Sie diese Kennung, um auf die Ressource zu verweisen: http://dx.doi.org/10.25673/101309
Titel: Laparoscopic augmented reality registration for oncological resection site repair
Autor(en): Joeres, FabianIn der Gemeinsamen Normdatei der DNB nachschlagen
Mielke, Tonia
Hansen, ChristianIn der Gemeinsamen Normdatei der DNB nachschlagen
Erscheinungsdatum: 2021
Art: Artikel
Sprache: Englisch
URN: urn:nbn:de:gbv:ma9:1-1981185920-1032646
Schlagwörter: Augmented reality
Laparoscopic surgery
Partial nephrectomy
Registration
Zusammenfassung: Purpose Resection site repair during laparoscopic oncological surgery (e.g. laparoscopic partial nephrectomy) poses some unique challenges and opportunities for augmented reality (AR) navigation support. This work introduces an AR registration workflow that addresses the time pressure that is present during resection site repair. Methods We propose a two-step registration process: the AR content is registered as accurately as possible prior to the tumour resection (the primary registration). This accurate registration is used to apply artificial fiducials to the physical organ and the virtual model. After the resection, these fiducials can be used for rapid re-registration (the secondary registration). We tested this pipeline in a simulated-use study with N = 18 participants. We compared the registration accuracy and speed for our method and for landmark-based registration as a reference. Results Acquisition of and, thereby, registration with the artificial fiducials were significantly faster than the initial use of anatomical landmarks. Our method also had a trend to be more accurate in cases in which the primary registration was successful. The accuracy loss between the elaborate primary registration and the rapid secondary registration could be quantified with a mean target registration error increase of 2.35mm. Conclusion This work introduces a registration pipeline for AR navigation support during laparoscopic resection site repair and provides a successful proof-of-concept evaluation thereof. Our results indicate that the concept is better suited than landmark-based registration during this phase, but further work is required to demonstrate clinical suitability and applicability.
URI: https://opendata.uni-halle.de//handle/1981185920/103264
http://dx.doi.org/10.25673/101309
Open-Access: Open-Access-Publikation
Nutzungslizenz: (CC BY 4.0) Creative Commons Namensnennung 4.0 International(CC BY 4.0) Creative Commons Namensnennung 4.0 International
Sponsor/Geldgeber: Projekt DEAL 2021
Journal Titel: International journal of computer assisted radiology and surgery
Verlag: Springer
Verlagsort: Berlin
Band: 16
Heft: 9
Originalveröffentlichung: 10.1007/s11548-021-02336-x
Seitenanfang: 1577
Seitenende: 1586
Enthalten in den Sammlungen:Fakultät für Informatik (OA)

Dateien zu dieser Ressource:
Datei Beschreibung GrößeFormat 
Joeres et al._ Laparoscopic augmented_2021.pdfZweitveröffentlichung1.38 MBAdobe PDFMiniaturbild
Öffnen/Anzeigen